Exfo FastReporter 2 User Manual Page 353

  • Download
  • Add to my manuals
  • Print
  • Page
    / 396
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 352
Description of Event Types
Post Processing Software 343
Non-Reflective Event
Non-Reflective Event
This event is characterized by a sudden decrease in the Rayleigh
backscatter signal level. It appears as a discontinuity in the downward
slope of the trace signal.
This event is often caused by splices, macrobends, or microbends in
the fiber.
A loss value is specified for non-reflective events. There is no
reflectance specified for this type of event.
If you set thresholds, the application indicates a non-reflective fault in
the events table, whenever a value exceeds the loss threshold.
Reflected power
(dB)
Distance
(km)
Non-reflective
event
Linear downward slope
due to Rayleigh
backscatter
Page view 352