Remote Control
Tunable Laser Source 89
General Commands
*TST?
Description This query initiates an internal self-test and returns a binary
value indicating the results of the test.
Syntax *TST?
Response An integer indicating the sum of all corresponding errors:
0: No errors
1: Interlock
2: Laser over-current
4: TEC failure
8: Flash not present
16: Flash time-out
32: EEprom not detected
64: EEprom check sum
128: Motor
256: Mechanic zero not found
512: Temperature
1024: FPGA
2048: Electronic power control
4096: Sweep
8192: Memory