Measuring Optical Return Loss
Multifunction Loss Tester 33
Performing and Saving ORL Measurements
5. Connect one end of a patchcord to the FASTEST SM port of your unit
leaving the other end unconnected.
6. Perform an ORL reference as follows:
Note: During the reference, the patchcord end should remain in the air;
reflections occurring at a fiber-to-air interface correspond to a constant of
14.7 dB.
Press ORL Reference.