Exfo OTDR 2 User Manual Page 181

  • Download
  • Add to my manuals
  • Print
  • Page
    / 219
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 180
OTDR 173
A Technical Specifications
MAX-710B
IMPORTANT
The following technical specifications can change without notice.
The information presented in this section is provided as a reference
only. To obtain this product’s most recent technical specifications,
visit the EXFO Web site at www.exfo.com.
TECHNICAL SPECIFICATIONS M
AX
T
ESTER
710B
Display 178 mm (7 in) outdoor-enhanced touchscreen, 800 x 480 TFT
Interfaces Two USB 2.0 ports
RJ-45 LAN 10/100 Mbit/s
Storage 2 GB internal memory (20 000 OTDR traces, typical)
Batteries Rechargeable lithium-polymer battery
12 hours of operation as per Telcordia (Bellcore) TR-NWT-001138
Power supply Power supply AC/DC adapter, input 100-240 VAC, 50-60 Hz, 9-16 V DCIN 15 Watts minimum
Wavelength (nm)
b
1310/1550
Dynamic range (dB)
c
30/28
Event dead zone (m)
d
1
Attenuation dead zone (m)
d
4
Distance range (km) 0.1 to 160 km
Pulse width (ns) 5 ns to 20 us
Linearity (dB/dB) ±0.05
Loss threshold (dB) 0.01
Loss resolution (dB) 0.001
Sampling resolution (m) 0.04 to 5
Sampling points Up to 256 000
Distance uncertainty (m)
e
±(0.75 + 0.005 % x distance + sampling resolution)
Measurement time User-defined (60 min. maximum)
Reflectance accuracy (dB) ±2
Typical real-time refresh (Hz) 3
Laser safety 1M
Notes
a. All specifications valid at 23 °C ± 2 °C with an FC/APC connector, unless otherwise specified.
b. Typical.
c. Typical dynamic range with longest pulse and three-minute averaging at SNR = 1.
d. Typical dead zone for reflectance below –55 dB, using a 5 ns pulse. Attenuation dead zone at 1310 nm is 5 m typical with reflectance below –45 dB.
e. Does not include uncertainty due to fiber index.
Page view 180
1 2 ... 176 177 178 179 180 181 182 183 184 185 186 ... 218 219

Comments to this Manuals

No comments