Exfo FTB-9100 for FTB-500 Optical Switch User Manual Page 51

  • Download
  • Add to my manuals
  • Print
  • Page
    / 72
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 50
SCPI Command Reference
Optical Switch 45
Product-Specific Commands—Description
:ROUTe[1..n]:SCAN
Description This command sets the switch to a specific
channel.
On *RST, the selected output channel depends
on the switch configuration:
a) For 1xN switch configuration: Output 1.
b) For 2x4 switch configuration: no continuity for
both Input 1 and Input 2.
c) For 2x2 switch configuration: BYPASS state.
Syntax :ROUTe[1..n]:SCAN<wsp><Position>
Parameter(s) Position:
The program data syntax for <Position> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
The <Position> is a number from 0 to 32,
depending on switch configuration as follows:
a) For a 1xN switch configuration, a number from
1 to 32, depending on the number of output
channels.
b) For a 2x4 switch, a number from 0 to 5, where
0, means no continuity for both Input 1 and
Input 2.
1, means no continuity for Input 1, and Input 2
is connected to Output 1.
Page view 50
1 2 ... 46 47 48 49 50 51 52 53 54 55 56 ... 71 72

Comments to this Manuals

No comments