SCPI Command Reference
50 FTB-9100
Product-Specific Commands—Description
:ROUTe[1..n]:SCAN:ADJust:AUTO
Description This command sets the <Optimize for
repeatability> option. When this option is
selected, the switch mechanism returns to a
reference position before aligning to a new
position. This ensures optimum repeatability.
When this option is not selected, the switch
mechanism goes directly to the new channel
position. This provides faster switching times, but
less repeatability.
On *RST, the value of this setting is OFF.
Syntax :ROUTe[1..n]:SCAN:ADJust:AUTO<wsp><Auto
Adjust>
Parameter(s) AutoAdjust:
The program data syntax for <AutoAdjust> is
defined as a <Boolean Program Data> element.
The <AutoAdjust> special forms ON and OFF
are accepted on input for increased readability.
ON corresponds to 1 and OFF corresponds to 0.
The <AutoAdjust> parameter enables or
disables the optimization for repeatability.
1 or ON, enables the optimization for
repeatability.
0 or OFF, disables the optimization for
repeatability.