Description of OTDR Event Types
Data Post-Processing Software 207
Continuous Fiber
Continuous Fiber
This event indicates that the selected acquisition range was shorter than
the fiber length.
³ The fiber end was not detected because the analysis process ended
before reaching the end of the fiber.
³ The acquisition distance range should therefore be increased to a
value greater than the fiber length.
³ There is no loss or reflectance specified for continuous fiber events.
Reflected power
(dB)
Continuous
fiber
Distance
(km)