Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT User Manual

Browse online or download User Manual for Measuring instruments Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT. EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434) User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 537
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
User Guide
Optical Spectrum Analyzer for FTB-500
FTB-5240S/5240BP
Page view 0
1 2 3 4 5 6 ... 536 537

Summary of Contents

Page 1 - FTB-5240S/5240BP

User GuideOptical Spectrum Analyzer for FTB-500FTB-5240S/5240BP

Page 2

Introducing the FTB-5240S/5240BP Optical Spectrum Analyzer2 FTB-5240S/5240BPModelsModelsThe Optical Spectrum Analyzer comes in different models:³ FTB-

Page 3 - Contents

Setting Up the Instrument in WDM Mode92 FTB-5240S/5240BPSetting Up WDM Analysis Parameters4. In the Save As window, enter a file name and press Save.

Page 4

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 93Setting Up WDM Analysis ParametersTo import a test setup:1. From the Main Menu, press

Page 5

Setting Up the Instrument in WDM Mode94 FTB-5240S/5240BPSetting Up WDM Analysis Parameters2. Select the Favorites tab.3. Press Import to import an ana

Page 6 - Certification Information

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 95Setting Up WDM Analysis Parameters5. If you want to apply the above settings to the a

Page 7 - DECLARATION OF CONFORMITY

Setting Up the Instrument in WDM Mode96 FTB-5240S/5240BPSetting Up WDM Analysis ParametersTo delete a test configuration:1. From the Main Menu, press

Page 8

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 97Setting Up WDM Analysis Parameters2. Select the Favorites tab.

Page 9 - Spectrum Analyzer

Setting Up the Instrument in WDM Mode98 FTB-5240S/5240BPSetting Up WDM Analysis Parameters3. To delete a configuration file from the favorites list, s

Page 10

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 99Setting Up Acquisition ParametersSetting Up Acquisition ParametersBefore performing y

Page 11 - Typical Applications

Setting Up the Instrument in WDM Mode100 FTB-5240S/5240BPSetting Up Acquisition ParametersTo set parameters in the Acquisition tab:1. From the main wi

Page 12

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 101Setting Up Acquisition Parameters³ If you are performing an Averaging or In-Band typ

Page 13 - 2 Safety Information

Introducing the FTB-5240S/5240BP Optical Spectrum AnalyzerOptical Spectrum Analyzer 3Typical ApplicationsTypical ApplicationsYou can use your Optical

Page 15 - Optical Spectrum Analyzer

Optical Spectrum Analyzer 1036 Setting Up the Instrument in Drift ModeBefore performing a spectral analysis in the Drift mode, you must set up your OS

Page 16 - 8 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode104 FTB-5240S/5240BP³ The easiest way to set up the instrument, especially when the operator does not know in a

Page 17 - FTB-500 left panel

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 105Defining PreferencesDefining PreferencesThe preferences window allows you to set g

Page 18 - 10 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode106 FTB-5240S/5240BPDefining Preferences2. Select the General tab.3. Define the general parameters as needed.4.

Page 19 - Optical Spectrum Analyzer 11

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 107Defining PreferencesTo enter link and location information:1. From the Main Menu,

Page 20 - Application

Setting Up the Instrument in Drift Mode108 FTB-5240S/5240BPDefining Preferences3. Under System and link information, define the following parameters a

Page 21

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 109Defining PreferencesTo enter comments:1. From the Main Menu, press Preferences.

Page 22 - Exiting the Application

Setting Up the Instrument in Drift Mode110 FTB-5240S/5240BPDefining Preferences2. Select the Comments tab. 3. Enter your comments for the current trac

Page 23 - Spectrum Analyzer for a Test

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 111Defining PreferencesDefining Display ParametersThe application allows you to set d

Page 24 - 16 FTB-5240S/5240BP

Introducing the FTB-5240S/5240BP Optical Spectrum Analyzer4 FTB-5240S/5240BPConventionsConventionsBefore using the product described in this manual, y

Page 25

Setting Up the Instrument in Drift Mode112 FTB-5240S/5240BPDefining Preferences2. Select the Display tab.

Page 26 - Selecting a Test Mode

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 113Defining Preferences3. Select the spectral unit you want to work with. With the na

Page 27 - To select a test mode:

Setting Up the Instrument in Drift Mode114 FTB-5240S/5240BPDefining Preferences4. Select the label that will appear on the peaks in the graph.

Page 28 - 20 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 115Defining PreferencesNote: The channel name and channel number cannot be shown at t

Page 29 - Nulling Electrical Offsets

Setting Up the Instrument in Drift Mode116 FTB-5240S/5240BPDefining Preferences5. Select Show to enable horizontal markers in the Markers tab. If Hide

Page 30 - 3. Press Nulling

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 117Defining Preferences6. Select the background color scheme for Graph as desired. 7.

Page 31

Setting Up the Instrument in Drift Mode118 FTB-5240S/5240BPDefining PreferencesCustomizing Drift Results TableIt is possible to select which results y

Page 32

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 119Defining Preferences2. Select the Drift Results tab. 3. Select which parameters yo

Page 33 - Performing User Calibration

Setting Up the Instrument in Drift Mode120 FTB-5240S/5240BPDefining PreferencesCustomizing Drift File NameThe application shall provide a way to defin

Page 34 - 26 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 121Defining Preferences2. Select the File Name tab. 3. Select which parameters you wa

Page 35

Optical Spectrum Analyzer 52 Safety InformationYour instrument is a Class 1 laser product in compliance with standards IEC 60825-1:2007 and 21 CFR 104

Page 36 - 28 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode122 FTB-5240S/5240BPDefining Preferences4. Press Up or Down to change the order in which the field values will

Page 37

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 123Setting Up Drift Analysis ParametersSetting Up Drift Analysis ParametersThis secti

Page 38 - 30 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode124 FTB-5240S/5240BPSetting Up Drift Analysis ParametersDefining General SettingsThe general analysis parameter

Page 39

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 125Setting Up Drift Analysis Parameters2. Select the General tab.

Page 40 - 32 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode126 FTB-5240S/5240BPSetting Up Drift Analysis Parameters3. Under Default channel settings, define the following

Page 41

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 127Setting Up Drift Analysis Parametersand be considered as one multi-peak signal. Wi

Page 42 - To save a user calibration:

Setting Up the Instrument in Drift Mode128 FTB-5240S/5240BPSetting Up Drift Analysis Parameters³ Noise for OSNR: indicates which calculation method to

Page 43

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 129Setting Up Drift Analysis Parameters³ Peak detection level (dBm): indicates the mi

Page 44 - 36 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode130 FTB-5240S/5240BPSetting Up Drift Analysis Parameters³ Power offset (dB): indicates the offset value applied

Page 45 - WDM Mode

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 131Setting Up Drift Analysis ParametersYou can set your pass/fail threshold limits in

Page 47

Setting Up the Instrument in Drift Mode132 FTB-5240S/5240BPSetting Up Drift Analysis ParametersDefining Channel ThresholdsThe thresholds will be appli

Page 48 - 40 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 133Setting Up Drift Analysis ParametersWhen thresholds are globally disabled, the res

Page 49

Setting Up the Instrument in Drift Mode134 FTB-5240S/5240BPSetting Up Drift Analysis Parameters2. Select the Thresholds tab.

Page 50

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 135Setting Up Drift Analysis Parameters3. Select the Activate all thresholds option,

Page 51

Setting Up the Instrument in Drift Mode136 FTB-5240S/5240BPSetting Up Drift Analysis Parameters4. Enter values in the boxes as explained below:³ Wavel

Page 52 - 44 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 137Setting Up Drift Analysis ParametersManaging ChannelsTesting DWDM systems involves

Page 53

Setting Up the Instrument in Drift Mode138 FTB-5240S/5240BPSetting Up Drift Analysis Parameters2. Select the Channels tab. 3. By default, the channel

Page 54

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 139Setting Up Drift Analysis Parameters³ Stop range (nm or Thz): indicates the ending

Page 55

Setting Up the Instrument in Drift Mode140 FTB-5240S/5240BPSetting Up Drift Analysis Parameters5. Press OK to return to the Channels window, which now

Page 56 - 48 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 141Setting Up Drift Analysis ParametersTo edit the parameters of a specific channel:1

Page 57

Optical Spectrum Analyzer 73 Getting Started with Your Optical Spectrum AnalyzerInserting and Removing Test ModulesTo insert a module into the FTB-500

Page 58 - 50 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode142 FTB-5240S/5240BPSetting Up Drift Analysis Parameters2. Select the Channels tab.

Page 59

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 143Setting Up Drift Analysis Parameters3. Select the channel or channels to be modifi

Page 60 - Customizing WDM Results Table

Setting Up the Instrument in Drift Mode144 FTB-5240S/5240BPSetting Up Drift Analysis Parameters5. Modify the settings as needed. For more information

Page 61 - Optical Spectrum Analyzer 53

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 145Setting Up Drift Analysis ParametersTo add current peaks:Note: You can add current

Page 62 - 54 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode146 FTB-5240S/5240BPSetting Up Drift Analysis Parameters2. Select the Channels tab. 3. Press Import from Trace.

Page 63 - Customizing WDM File Name

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 147Setting Up Drift Analysis ParametersManaging FavoritesFavorites are configurations

Page 64

Setting Up the Instrument in Drift Mode148 FTB-5240S/5240BPSetting Up Drift Analysis Parameters2. Select the Favorites tab.

Page 65

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 149Setting Up Drift Analysis Parameters3. To apply the settings from a favorite file

Page 66 - 58 FTB-5240S/5240BP

Setting Up the Instrument in Drift Mode150 FTB-5240S/5240BPSetting Up Drift Analysis ParametersTo save a test configuration:1. From the Main Menu, pre

Page 67

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 151Setting Up Drift Analysis Parameters2. Select the Favorites tab.

Page 68

Getting Started with Your Optical Spectrum Analyzer8 FTB-5240S/5240BPInserting and Removing Test Modules3. Take the module and place it so that the co

Page 69 - Optical Spectrum Analyzer 61

Setting Up the Instrument in Drift Mode152 FTB-5240S/5240BPSetting Up Drift Analysis Parameters3. To save an analysis setup to a file, press Save As.

Page 70 - 62 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 153Setting Up Drift Analysis Parameters4. In the Save As window, enter a file name an

Page 71 - Optical Spectrum Analyzer 63

Setting Up the Instrument in Drift Mode154 FTB-5240S/5240BPSetting Up Drift Analysis ParametersTo import a test setup:1. From the Main Menu, press Ana

Page 72

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 155Setting Up Drift Analysis Parameters2. Select the Favorites tab.3. Press Import to

Page 73 - Optical Spectrum Analyzer 65

Setting Up the Instrument in Drift Mode156 FTB-5240S/5240BPSetting Up Drift Analysis Parameters5. Press OK to load the configuration and close the win

Page 74 - 66 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 157Setting Up Drift Analysis Parameters2. Select the Favorites tab.

Page 75 - Defining Global Thresholds

Setting Up the Instrument in Drift Mode158 FTB-5240S/5240BPSetting Up Drift Analysis Parameters3. To delete a configuration file from the favorites li

Page 76 - To define global thresholds:

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 159Setting Up Acquisition ParametersSetting Up Acquisition ParametersBefore performin

Page 77 - Optical Spectrum Analyzer 69

Setting Up the Instrument in Drift Mode160 FTB-5240S/5240BPSetting Up Acquisition ParametersTo set parameters in the Acquisition tab:1. From the main

Page 78 - 70 FTB-5240S/5240BP

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 161Setting Up Acquisition Parameters3. Select the acquisition type. ³ If you are perf

Page 79 - Defining Default Thresholds

Getting Started with Your Optical Spectrum AnalyzerOptical Spectrum Analyzer 9Inserting and Removing Test Modules7. While applying slight pressure to

Page 80

Setting Up the Instrument in Drift Mode162 FTB-5240S/5240BPSetting Up Acquisition ParametersThe wavelength range covered within these bands of the spe

Page 81 - Optical Spectrum Analyzer 73

Setting Up the Instrument in Drift ModeOptical Spectrum Analyzer 163Setting Up Acquisition Parameters3. Select a sampling unit and enter a sampling co

Page 82

Setting Up the Instrument in Drift Mode164 FTB-5240S/5240BPSetting Up Acquisition Parameters7. Select the Keep historical traces in subfolder option t

Page 83 - Optical Spectrum Analyzer 75

Optical Spectrum Analyzer 1657 Setting Up the Instrument in DFB ModeBefore performing a spectral analysis in the DFB mode, you must set up your OSA mo

Page 84

Setting Up the Instrument in DFB Mode166 FTB-5240S/5240BPDefining PreferencesDefining PreferencesThe preferences window allows you to set general info

Page 85 - Optical Spectrum Analyzer 77

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 167Defining Preferences2. Select the General tab.3. Define the general parameters as ne

Page 86 - 78 FTB-5240S/5240BP

Setting Up the Instrument in DFB Mode168 FTB-5240S/5240BPDefining PreferencesTo enter link and location information:1. From the Main Menu, press Prefe

Page 87 - Optical Spectrum Analyzer 79

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 169Defining Preferences3. Under System and link information, define the following param

Page 88 - 4. Press Edit Selection

Setting Up the Instrument in DFB Mode170 FTB-5240S/5240BPDefining PreferencesTo enter comments:1. From the Main Menu, press Preferences.

Page 89 - Optical Spectrum Analyzer 81

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 171Defining Preferences2. Select the Comments tab. 3. Enter your comments for the curre

Page 90 - 82 FTB-5240S/5240BP

Getting Started with Your Optical Spectrum Analyzer10 FTB-5240S/5240BPInserting and Removing Test ModulesTo remove a module from the FTB-500:1. Exit T

Page 91

Setting Up the Instrument in DFB Mode172 FTB-5240S/5240BPDefining PreferencesDefining Display ParametersThe application allows you to set display sett

Page 92 - 84 FTB-5240S/5240BP

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 173Defining Preferences2. Select the Display tab.

Page 93 - Optical Spectrum Analyzer 85

Setting Up the Instrument in DFB Mode174 FTB-5240S/5240BPDefining Preferences3. Select the spectral unit you want to work with. With the nanometer (nm

Page 94

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 175Defining Preferences4. Select the background color scheme for Graph as desired. 5. P

Page 95

Setting Up the Instrument in DFB Mode176 FTB-5240S/5240BPDefining PreferencesCustomizing DFB File NameThe application shall provide a way to define th

Page 96 - 88 FTB-5240S/5240BP

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 177Defining Preferences2. Select the File Name tab. 3. Select which parameters you want

Page 97 - Optical Spectrum Analyzer 89

Setting Up the Instrument in DFB Mode178 FTB-5240S/5240BPDefining Preferences4. Press Up or Down to change the order in which the field values will ap

Page 98

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 179Setting Up Acquisition ParametersSetting Up Acquisition ParametersBefore performing

Page 99

Setting Up the Instrument in DFB Mode180 FTB-5240S/5240BPSetting Up Acquisition ParametersTo set parameters in the Acquisition tab:1. From the main wi

Page 100 - 92 FTB-5240S/5240BP

Setting Up the Instrument in DFB ModeOptical Spectrum Analyzer 181Setting Up Acquisition Parameters³ If you are performing an Averaging type acquisiti

Page 101 - To import a test setup:

Getting Started with Your Optical Spectrum AnalyzerOptical Spectrum Analyzer 11Inserting and Removing Test Modules5. Hold the module by its sides or b

Page 103 - Optical Spectrum Analyzer 95

Optical Spectrum Analyzer 1838 Setting Up the Instrument in Spectral Transmittance ModeBefore performing a spectral analysis in the Spectral Transmitt

Page 104 - 96 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance Mode184 FTB-5240S/5240BPDefining PreferencesDefining PreferencesThe preferences window allows you

Page 105 - 2. Select the Favorites tab

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 185Defining Preferences2. Select the General tab.3. Define the gener

Page 106 - 98 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance Mode186 FTB-5240S/5240BPDefining PreferencesTo enter link and location information:1. From the Mai

Page 107 - Optical Spectrum Analyzer 99

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 187Defining Preferences3. Under System and link information, define

Page 108 - 100 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance Mode188 FTB-5240S/5240BPDefining PreferencesTo enter comments:1. From the Main Menu, press Prefere

Page 109 - Optical Spectrum Analyzer 101

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 189Defining Preferences2. Select the Comments tab. 3. Enter your com

Page 110

Setting Up the Instrument in Spectral Transmittance Mode190 FTB-5240S/5240BPDefining PreferencesDefining Display ParametersThe application allows you

Page 111 - Drift Mode

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 191Defining Preferences2. Select the Display tab.

Page 112 - 104 FTB-5240S/5240BP

ii FTB-5240S/5240BPCopyright © 2009–2010 EXFO Inc. All rights reserved. No part of this publication may be reproduced, stored in a retrieval system or

Page 113 - Defining Preferences

Getting Started with Your Optical Spectrum Analyzer12 FTB-5240S/5240BPStarting the Optical Spectrum Analyzer ApplicationStarting the Optical Spectrum

Page 114 - 106 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance Mode192 FTB-5240S/5240BPDefining Preferences3. Select the spectral unit you want to work with. Wit

Page 115

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 193Defining Preferences4. Select Show to enable horizontal markers i

Page 116 - IMPORTANT

Setting Up the Instrument in Spectral Transmittance Mode194 FTB-5240S/5240BPDefining Preferences5. Select the background color scheme for Graph as des

Page 117 - To enter comments:

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 195Defining PreferencesCustomizing Spectral Transmittance File NameT

Page 118 - 110 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance Mode196 FTB-5240S/5240BPDefining Preferences2. Select the File Name tab. 3. Select which parameter

Page 119 - Defining Display Parameters

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 197Defining Preferences4. Press Up or Down to change the order in wh

Page 120 - 2. Select the Display tab

Setting Up the Instrument in Spectral Transmittance Mode198 FTB-5240S/5240BPSetting Up Spectral Transmittance Analysis ParametersSetting Up Spectral T

Page 121

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 199Setting Up Spectral Transmittance Analysis ParametersDefining ST

Page 122 - 114 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance Mode200 FTB-5240S/5240BPSetting Up Spectral Transmittance Analysis Parameters2. Select the ST Anal

Page 123

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 201Setting Up Spectral Transmittance Analysis Parameters3. Under Glo

Page 124 - 116 FTB-5240S/5240BP

Getting Started with Your Optical Spectrum AnalyzerOptical Spectrum Analyzer 13Starting the Optical Spectrum Analyzer ApplicationThe main window (show

Page 125

Setting Up the Instrument in Spectral Transmittance Mode202 FTB-5240S/5240BPSetting Up Spectral Transmittance Analysis Parameters³ Channel distance (G

Page 126 - 118 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 203Setting Up Spectral Transmittance Analysis ParametersTo edit the

Page 127

Setting Up the Instrument in Spectral Transmittance Mode204 FTB-5240S/5240BPSetting Up Spectral Transmittance Analysis Parameters4. If you want to app

Page 128 - Customizing Drift File Name

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 205Setting Up Acquisition ParametersSetting Up Acquisition Parameter

Page 129 - Optical Spectrum Analyzer 121

Setting Up the Instrument in Spectral Transmittance Mode206 FTB-5240S/5240BPSetting Up Acquisition ParametersTo set parameters in the Acquisition tab:

Page 130 - 122 FTB-5240S/5240BP

Setting Up the Instrument in Spectral Transmittance ModeOptical Spectrum Analyzer 207Setting Up Acquisition Parameters³ If you are performing an Avera

Page 131 - Optical Spectrum Analyzer 123

Setting Up the Instrument in Spectral Transmittance Mode208 FTB-5240S/5240BPSetting Up Acquisition ParametersThe wavelength range covered within these

Page 132

Optical Spectrum Analyzer 2099 Setting Up the Instrument in EDFA ModeBefore performing a spectral analysis in the EDFA mode, you must set up your OSA

Page 133 - 2. Select the General tab

Setting Up the Instrument in EDFA Mode210 FTB-5240S/5240BPDefining PreferencesDefining PreferencesThe preferences window allows you to set general inf

Page 134 - 126 FTB-5240S/5240BP

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 211Defining Preferences2. Select the General tab.3. Define the general parameters as n

Page 135 - Optical Spectrum Analyzer 127

Getting Started with Your Optical Spectrum Analyzer14 FTB-5240S/5240BPExiting the ApplicationStatus BarThe status bar, located at the bottom of the ma

Page 136 - 128 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode212 FTB-5240S/5240BPDefining PreferencesTo enter link and location information:1. From the Main Menu, press Pref

Page 137 - Optical Spectrum Analyzer 129

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 213Defining Preferences2. Select the Information tab.3. Under System and link informat

Page 138 - press Edit % button

Setting Up the Instrument in EDFA Mode214 FTB-5240S/5240BPDefining Preferences4. Under Location Information, define the following parameters as needed

Page 139 - Threshold Limit Definition

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 215Defining Preferences2. Select the Comments tab. 3. Enter your comments for the curr

Page 140 - Defining Channel Thresholds

Setting Up the Instrument in EDFA Mode216 FTB-5240S/5240BPDefining PreferencesDefining Display ParametersThe application allows you to set display set

Page 141 - Optical Spectrum Analyzer 133

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 217Defining Preferences2. Select the Display tab.

Page 142 - 134 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode218 FTB-5240S/5240BPDefining Preferences3. Select the spectral unit you want to work with. With the nanometer (n

Page 143 - Optical Spectrum Analyzer 135

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 219Defining Preferences4. Select the label that will appear on the peaks in the graph.

Page 144 - 136 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode220 FTB-5240S/5240BPDefining PreferencesNote: The channel name and channel number cannot be shown at the same ti

Page 145 - Managing Channels

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 221Defining Preferences5. Select Show to display the empty channels from the channel l

Page 146 - 138 FTB-5240S/5240BP

Optical Spectrum Analyzer 154 Preparing Your Optical Spectrum Analyzer for a TestCleaning and Connecting Optical FibersTo connect the fiber-optic cabl

Page 147 - Optical Spectrum Analyzer 139

Setting Up the Instrument in EDFA Mode222 FTB-5240S/5240BPDefining Preferences6. Select Show to enable horizontal markers in the Markers tab. If Hide

Page 148 - 140 FTB-5240S/5240BP

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 223Defining Preferences7. Select the background color scheme for Graph as desired. 8.

Page 149 - Optical Spectrum Analyzer 141

Setting Up the Instrument in EDFA Mode224 FTB-5240S/5240BPDefining PreferencesCustomizing EDFA Results TableIt is possible to select which results you

Page 150 - 2. Select the Channels tab

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 225Defining Preferences2. Select the EDFA Results tab. 3. Select which parameters you

Page 151 - Optical Spectrum Analyzer 143

Setting Up the Instrument in EDFA Mode226 FTB-5240S/5240BPDefining Preferences³ Gain: indicates the gain (Output Signal Power - Input Signal Power) fo

Page 152 - 144 FTB-5240S/5240BP

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 227Defining PreferencesCustomizing EDFA File NameThe application shall provide a way t

Page 153 - To add current peaks:

Setting Up the Instrument in EDFA Mode228 FTB-5240S/5240BPDefining Preferences2. Select the File Name tab. 3. Select which parameters you want to incl

Page 154 - 146 FTB-5240S/5240BP

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 229Defining Preferences4. Press Up or Down to change the order in which the field valu

Page 155 - Managing Favorites

Setting Up the Instrument in EDFA Mode230 FTB-5240S/5240BPSetting Up EDFA Analysis ParametersSetting Up EDFA Analysis ParametersThis section presents

Page 156

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 231Setting Up EDFA Analysis ParametersDefining General SettingsThe general analysis pa

Page 157 - Optical Spectrum Analyzer 149

Preparing Your Optical Spectrum Analyzer for a Test16 FTB-5240S/5240BPCleaning and Connecting Optical Fibers3. Carefully align the connector and port

Page 158 - To save a test configuration:

Setting Up the Instrument in EDFA Mode232 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters2. Select the General tab.

Page 159

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 233Setting Up EDFA Analysis Parameters3. Under Default channel settings, define the fo

Page 160 - 152 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode234 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters³ Channel width (GHz or nm): indicates the limit inside w

Page 161 - Optical Spectrum Analyzer 153

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 235Setting Up EDFA Analysis Parameters³ Peak detection level (dBm): indicates the mini

Page 162

Setting Up the Instrument in EDFA Mode236 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters³ Output wavelength offset (nm): indicates the offset val

Page 163 - Optical Spectrum Analyzer 155

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 237Setting Up EDFA Analysis Parameters5. If you want to apply the above settings to th

Page 164 - 156 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode238 FTB-5240S/5240BPSetting Up EDFA Analysis ParametersManaging ChannelsTesting DWDM systems involves characteri

Page 165

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 239Setting Up EDFA Analysis Parameters2. Select the Channels tab. 3. By default, the c

Page 166 - 158 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode240 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters4. Enter values in the boxes as explained below:³ Start r

Page 167 - Optical Spectrum Analyzer 159

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 241Setting Up EDFA Analysis Parameters³ Starting Value: Sets the increment starting va

Page 168 - 160 FTB-5240S/5240BP

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 17Installing the EXFO Universal Interface (EUI)Installing the EXFO Univer

Page 169 - Optical Spectrum Analyzer 161

Setting Up the Instrument in EDFA Mode242 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters7. Press OK to save the changes and close the window, or

Page 170

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 243Setting Up EDFA Analysis Parameters2. Select the Channels tab.

Page 171 - Optical Spectrum Analyzer 163

Setting Up the Instrument in EDFA Mode244 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters3. Select the channel or channels to be modified in the c

Page 172 - 164 FTB-5240S/5240BP

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 245Setting Up EDFA Analysis Parameters5. Modify the settings as needed. For more infor

Page 173 - DFB Mode

Setting Up the Instrument in EDFA Mode246 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters7. If you want to apply the above settings to the active

Page 174

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 247Setting Up EDFA Analysis ParametersTo add current peaks:Note: You can add current p

Page 175

Setting Up the Instrument in EDFA Mode248 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters3. Press Import from Trace. All peaks from the current tr

Page 176 - 168 FTB-5240S/5240BP

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 249Setting Up EDFA Analysis Parameters4. If you want to apply the above settings to th

Page 177

Setting Up the Instrument in EDFA Mode250 FTB-5240S/5240BPSetting Up EDFA Analysis ParametersManaging FavoritesFavorites are configurations files that

Page 178

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 251Setting Up EDFA Analysis Parameters2. Select the Favorites tab.

Page 179

Preparing Your Optical Spectrum Analyzer for a Test18 FTB-5240S/5240BPSelecting a Test ModeSelecting a Test ModeYour Optical Spectrum Analyzer gives y

Page 180

Setting Up the Instrument in EDFA Mode252 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters3. To apply the settings from a favorite file to the curr

Page 181

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 253Setting Up EDFA Analysis Parameters4. If you want to apply the above settings to th

Page 182 - 174 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode254 FTB-5240S/5240BPSetting Up EDFA Analysis ParametersTo save a test configuration:1. From the Main Menu, press

Page 183

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 255Setting Up EDFA Analysis Parameters2. Select the Favorites tab.

Page 184 - Customizing DFB File Name

Setting Up the Instrument in EDFA Mode256 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters3. To save an analysis setup to a file, press Save As. Th

Page 185 - Optical Spectrum Analyzer 177

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 257Setting Up EDFA Analysis Parameters5. If you want to apply the above settings to th

Page 186 - 178 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode258 FTB-5240S/5240BPSetting Up EDFA Analysis ParametersTo import a test setup:1. From the Main Menu, press Analy

Page 187 - Optical Spectrum Analyzer 179

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 259Setting Up EDFA Analysis Parameters2. Select the Favorites tab.3. Press Import to i

Page 188 - 180 FTB-5240S/5240BP

Setting Up the Instrument in EDFA Mode260 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters4. From the Import window, select the file you want to im

Page 189 - Optical Spectrum Analyzer 181

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 261Setting Up EDFA Analysis Parameters5. If you want to apply the above settings to th

Page 190

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 19Selecting a Test ModeTo select a test mode:1. From the main menu, press

Page 191 - Spectral Transmittance Mode

Setting Up the Instrument in EDFA Mode262 FTB-5240S/5240BPSetting Up EDFA Analysis ParametersTo delete a test configuration:1. From the Main Menu, pre

Page 192

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 263Setting Up EDFA Analysis Parameters2. Select the Favorites tab.

Page 193

Setting Up the Instrument in EDFA Mode264 FTB-5240S/5240BPSetting Up EDFA Analysis Parameters3. To delete a configuration file from the favorites list

Page 194 - 186 FTB-5240S/5240BP

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 265Setting Up Acquisition ParametersSetting Up Acquisition ParametersBefore performing

Page 195

Setting Up the Instrument in EDFA Mode266 FTB-5240S/5240BPSetting Up Acquisition ParametersTo set parameters in the Acquisition tab:1. From the main w

Page 196

Setting Up the Instrument in EDFA ModeOptical Spectrum Analyzer 267Setting Up Acquisition Parameters³ If you are performing an Averaging type acquisit

Page 197

Setting Up the Instrument in EDFA Mode268 FTB-5240S/5240BPSetting Up Acquisition ParametersThe wavelength range covered within these bands of the spec

Page 198

Optical Spectrum Analyzer 26910 Testing DWDM SystemsOptical spectrum analysis is the measurement of optical power as a function of wavelength or frequ

Page 199

Testing DWDM Systems270 FTB-5240S/5240BPUsing the Discover Featurethan 0.001 nm or 0.001 GHz, then their width is reduce to the lower width. If the wi

Page 200 - 192 FTB-5240S/5240BP

Testing DWDM SystemsOptical Spectrum Analyzer 271Using the Discover FeatureTo start an automatic setup scan:From the Main Menu, press Discover.WDM Mod

Page 201

Preparing Your Optical Spectrum Analyzer for a Test20 FTB-5240S/5240BPSelecting a Test Mode2. Select the desired test mode.Once you select the mode, y

Page 202 - 194 FTB-5240S/5240BP

Testing DWDM Systems272 FTB-5240S/5240BPUsing the Discover FeatureNote: If you already have an active trace on screen that was modified, you will be p

Page 203

Testing DWDM SystemsOptical Spectrum Analyzer 273Starting a MeasurementStarting a MeasurementNote: Before starting a measurement you must select and c

Page 204

Testing DWDM Systems274 FTB-5240S/5240BPStarting a MeasurementDrift ModeDFB Mode

Page 205

Testing DWDM SystemsOptical Spectrum Analyzer 275Starting a MeasurementSpectral TransmittanceModeEDFA Mode

Page 206 - Parameters

Testing DWDM Systems276 FTB-5240S/5240BPStarting a MeasurementWhen an acquisition is undergoing in the module, the Start button changes to Stop. Press

Page 207 - Defining ST Analysis Settings

Optical Spectrum Analyzer 27711 Managing ResultsYour Optical Spectrum Analyzer allows you to manage results in the WDM, Drift, DFB, Spectral Transmitt

Page 208 - 200 FTB-5240S/5240BP

Managing Results278 FTB-5240S/5240BPManaging WDM Test ResultsManaging WDM Test ResultsThe application allows you to view and manage your WDM test resu

Page 209 - Optical Spectrum Analyzer 201

Managing ResultsOptical Spectrum Analyzer 279Managing WDM Test ResultsWhen the acquisition is taken (see Starting a Measurement on page 273 for detail

Page 210

Managing Results280 FTB-5240S/5240BPManaging WDM Test ResultsYour application allows you to toggle the view of your WDM main window. You can change th

Page 211 - Optical Spectrum Analyzer 203

Managing ResultsOptical Spectrum Analyzer 281Managing WDM Test ResultsFor 100 % lower tabs view, press .The graph will display peak indicators for al

Page 212 - 204 FTB-5240S/5240BP

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 21Nulling Electrical OffsetsNulling Electrical OffsetsThe offset nulling

Page 213 - Optical Spectrum Analyzer 205

Managing Results282 FTB-5240S/5240BPManaging WDM Test ResultsThe graph will also display the noise level for a channel with a doted line under the sel

Page 214 - 206 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 283Managing WDM Test ResultsResults TabThe results table shows channel results for the active trace. The res

Page 215 - Optical Spectrum Analyzer 207

Managing Results284 FTB-5240S/5240BPManaging WDM Test Results³ Noise (dBm): indicates the level of the noise for the selected channel (according to th

Page 216

Managing ResultsOptical Spectrum Analyzer 285Managing WDM Test ResultsMarkers TabYou can use markers to perform manual measurements and verification d

Page 217 - EDFA Mode

Managing Results286 FTB-5240S/5240BPManaging WDM Test ResultsTo define a marker:1. From the main window, press the Markers tab.2. Set the marker by en

Page 218

Managing ResultsOptical Spectrum Analyzer 287Managing WDM Test ResultsYou can also move the markers directly on the Graph tab. Drag the marker to the

Page 219

Managing Results288 FTB-5240S/5240BPManaging WDM Test ResultsTrace Info. TabAfter acquiring a trace, you might want to view details about the acquisit

Page 220 - 212 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 289Managing WDM Test Results3. Select the General tab.4. Edit the general information as required.Note: If y

Page 221

Managing Results290 FTB-5240S/5240BPManaging WDM Test ResultsTo edit trace information:1. From the main window, select the Trace Info. tab.2. Press Tr

Page 222 - 214 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 291Managing WDM Test ResultsTo edit comments:1. From the main window, select the Trace Info. tab.2. Press Tr

Page 223

ContentsOptical Spectrum Analyzer iiiContentsCertification Information ...

Page 224

Preparing Your Optical Spectrum Analyzer for a Test22 FTB-5240S/5240BPNulling Electrical Offsets3. Press Nulling. You are notified that the nulling is

Page 225

Managing Results292 FTB-5240S/5240BPManaging WDM Test ResultsChannel Results TabWhen you select a row from the Results tab, the Channel Results tab wi

Page 226 - 218 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 293Managing WDM Test Results2. Select a row from the Results tab to view the channel results for the selecte

Page 227

Managing Results294 FTB-5240S/5240BPManaging WDM Test Results³ Bandwidth 3.00 dB: indicates bandwidth measured by taking the width of a signal at 50 %

Page 228 - 220 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 295Managing WDM Test Results³ Noise for OSNR: indicates which calculation method to use for OSNR value. Fixe

Page 229

Managing Results296 FTB-5240S/5240BPManaging WDM Test ResultsYou can include selected channels results in the global results. These channels will be i

Page 230 - 222 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 297Managing WDM Test Results2. Select a row from the Results tab to view the channel results for the selecte

Page 231

Managing Results298 FTB-5240S/5240BPManaging WDM Test Results3. Select the Include in calculation option.

Page 232 - 224 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 299Managing WDM Test ResultsTo re-analyze channel results:1. From the main window, select the Channel Result

Page 233 - Optical Spectrum Analyzer 225

Managing Results300 FTB-5240S/5240BPManaging WDM Test Results3. Edit the values as explained below:³ Channel number: indicates the number of the chann

Page 234 - 226 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 301Managing WDM Test Results³ Signal power calculation: indicates which calculation method to apply for sign

Page 235 - Customizing EDFA File Name

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 23Performing User CalibrationPerforming User CalibrationCalibrating your

Page 236

Managing Results302 FTB-5240S/5240BPManaging WDM Test Results4. Edit the thresholds as explained below:³ Wavelength/Frequency (nm/GHz): indicates the

Page 237

Managing ResultsOptical Spectrum Analyzer 303Managing WDM Test ResultsNote: If a channel is already included in the channel list, the application will

Page 238 - 230 FTB-5240S/5240BP

Managing Results304 FTB-5240S/5240BPManaging WDM Test ResultsGlobal Results TabThe application allows you to view the global results of the current me

Page 239

Managing ResultsOptical Spectrum Analyzer 305Managing WDM Test Results³ Total power in scan range: indicates the total power for the selected scan ran

Page 240

Managing Results306 FTB-5240S/5240BPManaging WDM Test ResultsChanging Active Trace Analysis Parameters and Re-analyzingThe global parameters modifies

Page 241 - Optical Spectrum Analyzer 233

Managing ResultsOptical Spectrum Analyzer 307Managing WDM Test Results2. Press Global Parameters. 3. Select the Analysis tab. 4. Enter values in the b

Page 242 - 234 FTB-5240S/5240BP

Managing Results308 FTB-5240S/5240BPManaging WDM Test Results³ Wavelength offset (nm): indicates the offset value applied on the wavelength. You can u

Page 243 - Optical Spectrum Analyzer 235

Managing ResultsOptical Spectrum Analyzer 309Managing WDM Test ResultsChanging Default Channel ParametersDefault channel settings will define how any

Page 244 - 236 FTB-5240S/5240BP

Managing Results310 FTB-5240S/5240BPManaging WDM Test Results2. Press Global Parameters. 3. Select the Default Channel tab.

Page 245 - Optical Spectrum Analyzer 237

Managing ResultsOptical Spectrum Analyzer 311Managing WDM Test Results4. Enter values in the boxes as explained below:³ Clear the Activate default cha

Page 246

Preparing Your Optical Spectrum Analyzer for a Test24 FTB-5240S/5240BPPerforming User CalibrationA high level sequence of steps to perform user calibr

Page 247

Managing Results312 FTB-5240S/5240BPManaging WDM Test Results³ Signal power calculation: indicates which calculation method to apply for signal power

Page 248

Managing ResultsOptical Spectrum Analyzer 313Managing WDM Test Results5. Press OK to save the changes, re-analyze data and close the window, or press

Page 249 - Optical Spectrum Analyzer 241

Managing Results314 FTB-5240S/5240BPManaging WDM Test ResultsTo change global threshold parameters:1. From the main window, select the Global Results

Page 250 - 242 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 315Managing WDM Test Results2. Press Global Parameters. 3. Select the Global Thresholds tab. 4. Select the A

Page 251

Managing Results316 FTB-5240S/5240BPManaging WDM Test Results³ Average OSNR (dB): indicates the sum of the entire OSNR of the peaks detected in the cu

Page 252 - 244 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 317Managing WDM Test ResultsChanging Default Thresholds ParametersThe default thresholds will be applied to

Page 253 - Optical Spectrum Analyzer 245

Managing Results318 FTB-5240S/5240BPManaging WDM Test Results2. Press Global Parameters. 3. Select the Default Thresholds tab.

Page 254 - 246 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 319Managing WDM Test Results4. Enter values in the boxes as explained below:³ Wavelength/Frequency (nm/GHz):

Page 255 - Optical Spectrum Analyzer 247

Managing Results320 FTB-5240S/5240BPManaging Drift Test ResultsManaging Drift Test ResultsThe application allows you to view and manage your drift tes

Page 256 - 248 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 321Managing Drift Test ResultsYou can select a channel directly from the dashboard or from the Channel Histo

Page 257 - Optical Spectrum Analyzer 249

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 25Performing User CalibrationAn example of the calibration file format is

Page 258

Managing Results322 FTB-5240S/5240BPManaging Drift Test Results³ The dashboard shows a global status (all channels) for each parameter. This global st

Page 259

Managing ResultsOptical Spectrum Analyzer 323Managing Drift Test Results³ The dashboard displays a channel status (all parameters) for a given channel

Page 260 - 252 FTB-5240S/5240BP

Managing Results324 FTB-5240S/5240BPManaging Drift Test ResultsChannel Graph TabThe Channel Graph tab displays three different graphs for the selected

Page 261 - Optical Spectrum Analyzer 253

Managing ResultsOptical Spectrum Analyzer 325Managing Drift Test ResultsChannel History TabThe Channel History table shows channel results for the act

Page 262

Managing Results326 FTB-5240S/5240BPManaging Drift Test ResultsTo view channel history results:From the main window, select the Channel History tab.Re

Page 263

Managing ResultsOptical Spectrum Analyzer 327Managing Drift Test ResultsA small red marker ( ) will point down at the peak in the WDM Graph tab when y

Page 264 - 256 FTB-5240S/5240BP

Managing Results328 FTB-5240S/5240BPManaging Drift Test ResultsWDM Graph TabThe WDM Graph tab allows you to view the spectrum of the active trace for

Page 265 - Optical Spectrum Analyzer 257

Managing ResultsOptical Spectrum Analyzer 329Managing Drift Test ResultsWhen the acquisition is taken (see Starting a Measurement on page 273 for deta

Page 266

Managing Results330 FTB-5240S/5240BPManaging Drift Test ResultsMarkers TabYou can use markers to perform manual measurements and verification directly

Page 267

Managing ResultsOptical Spectrum Analyzer 331Managing Drift Test ResultsTo define a marker:1. From the main window, press the Markers tab.2. Set the m

Page 268 - Favorites window

Preparing Your Optical Spectrum Analyzer for a Test26 FTB-5240S/5240BPPerforming User Calibration2. Select the Calibration tab. Note: You cannot edit

Page 269 - Optical Spectrum Analyzer 261

Managing Results332 FTB-5240S/5240BPManaging Drift Test ResultsYou can also move the markers directly on the WDM Graph tab. Drag the marker to the des

Page 270 - 262 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 333Managing Drift Test ResultsTo edit general information:1. From the main window, select the Trace Info. ta

Page 271

Managing Results334 FTB-5240S/5240BPManaging Drift Test ResultsTo edit trace information:1. From the main window, select the Trace Info. tab.2. Press

Page 272 - 264 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 335Managing Drift Test ResultsTo edit comments:1. From the main window, select the Trace Info. tab.2. Press

Page 273 - Optical Spectrum Analyzer 265

Managing Results336 FTB-5240S/5240BPManaging Drift Test ResultsChannel Results TabWhen you select a channel in the Channel History tab, the Channel Re

Page 274 - 266 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 337Managing Drift Test Results2. Select a channel from the Channel History tab to view the channel results f

Page 275 - Optical Spectrum Analyzer 267

Managing Results338 FTB-5240S/5240BPManaging Drift Test Results³ Bandwidth x dB: indicates bandwidth measured by taking the width of a signal at x dB

Page 276

Managing ResultsOptical Spectrum Analyzer 339Managing Drift Test ResultsIn-Band narrow filter (InB nf): The In-Band narrow filter method uses the In-B

Page 277 - 10 Testing DWDM Systems

Managing Results340 FTB-5240S/5240BPManaging DFB Test ResultsManaging DFB Test ResultsThe application allows you to view and manage your DFB test resu

Page 278 - Using the Discover Feature

Managing ResultsOptical Spectrum Analyzer 341Managing DFB Test ResultsWhen the acquisition is taken (see Starting a Measurement on page 273 for detail

Page 279 - Testing DWDM Systems

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 27Performing User Calibration3. If a user calibration file exists, press

Page 280 - 272 FTB-5240S/5240BP

Managing Results342 FTB-5240S/5240BPManaging DFB Test ResultsResults TabYou can view the analysis of the DFB laser source from the Results tab.To v ie

Page 281 - Starting a Measurement

Managing ResultsOptical Spectrum Analyzer 343Managing DFB Test Results³ Right stopband: indicates the spectral position difference between the main mo

Page 282

Managing Results344 FTB-5240S/5240BPManaging DFB Test ResultsMarkers TabYou can use markers to perform manual measurements and verification directly o

Page 283

Managing ResultsOptical Spectrum Analyzer 345Managing DFB Test ResultsTo define a marker:1. From the main window, press the Markers tab.2. Set the mar

Page 284 - 276 FTB-5240S/5240BP

Managing Results346 FTB-5240S/5240BPManaging DFB Test ResultsNote: After using the zoom tools in the DFB Graph tab, you can only move the markers agai

Page 285 - 11 Managing Results

Managing ResultsOptical Spectrum Analyzer 347Managing DFB Test Results3. Select the General tab.4. Edit the general information as required.Note: If y

Page 286 - Managing WDM Test Results

Managing Results348 FTB-5240S/5240BPManaging DFB Test ResultsTo edit trace information:1. From the main window, select the Trace Info. tab.2. Press Tr

Page 287

Managing ResultsOptical Spectrum Analyzer 349Managing DFB Test ResultsTo edit comments:1. From the main window, select the Trace Info. tab.2. Press Tr

Page 288 - Managing Results

Managing Results350 FTB-5240S/5240BPManaging Spectral Transmittance Test ResultsManaging Spectral Transmittance Test ResultsThe application allows you

Page 289

Managing ResultsOptical Spectrum Analyzer 351Managing Spectral Transmittance Test ResultsWhen the acquisition is taken (see Starting a Measurement on

Page 290

Preparing Your Optical Spectrum Analyzer for a Test28 FTB-5240S/5240BPPerforming User CalibrationThe calibration values will replace the Correction fa

Page 291 - Results Tab

Managing Results352 FTB-5240S/5240BPManaging Spectral Transmittance Test ResultsResults TabThe results table shows the spectral transmittance results

Page 292 - 284 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 353Managing Spectral Transmittance Test ResultsMarkers TabYou can use markers to perform manual measurements

Page 293 - Markers Tab

Managing Results354 FTB-5240S/5240BPManaging Spectral Transmittance Test ResultsTo define a marker:1. From the main window, press the Markers tab.2. S

Page 294

Managing ResultsOptical Spectrum Analyzer 355Managing Spectral Transmittance Test ResultsYou can also move the markers directly on the ST Graph tab. D

Page 295

Managing Results356 FTB-5240S/5240BPManaging Spectral Transmittance Test ResultsTrace Info. TabAfter acquiring a trace, you might want to view details

Page 296 - Trace Info. Tab

Managing ResultsOptical Spectrum Analyzer 357Managing Spectral Transmittance Test Results3. Select the General tab.4. Edit the general information as

Page 297

Managing Results358 FTB-5240S/5240BPManaging Spectral Transmittance Test ResultsTo edit trace information:1. From the main window, select the Trace In

Page 298 - 290 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 359Managing Spectral Transmittance Test ResultsTo edit comments:1. From the main window, select the Trace In

Page 299

Managing Results360 FTB-5240S/5240BPManaging Spectral Transmittance Test ResultsST Results TabThe ST Results tab will show complete information about

Page 300 - Channel Results Tab

Managing ResultsOptical Spectrum Analyzer 361Managing Spectral Transmittance Test Results³ Insertion loss max (dB): indicates the maximum difference b

Page 301 - Optical Spectrum Analyzer 293

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 29Performing User Calibration5. Press Write to Module to apply the modifi

Page 302

Managing Results362 FTB-5240S/5240BPManaging Spectral Transmittance Test Results³ Input wavelength offset (nm): indicates the offset value applied on

Page 303

Managing ResultsOptical Spectrum Analyzer 363Managing Spectral Transmittance Test ResultsTo edit ST analysis parameters:1. From the main window, selec

Page 304 - 296 FTB-5240S/5240BP

Managing Results364 FTB-5240S/5240BPManaging Spectral Transmittance Test Results2. Press Global Parameters.

Page 305

Managing ResultsOptical Spectrum Analyzer 365Managing Spectral Transmittance Test Results3. Edit the values as explained below:³ Channel definition (G

Page 306

Managing Results366 FTB-5240S/5240BPManaging Spectral Transmittance Test Results³ Channel range (GHz or nm): indicates the limit inside which the powe

Page 307 - 2. Press Channel Parameters

Managing ResultsOptical Spectrum Analyzer 367Managing Spectral Transmittance Test Results³ Output wavelength offset (nm): indicates the offset value a

Page 308

Managing Results368 FTB-5240S/5240BPManaging EDFA Test ResultsManaging EDFA Test ResultsThe application allows you to view and manage your EDFA test r

Page 309

Managing ResultsOptical Spectrum Analyzer 369Managing EDFA Test ResultsWhen the acquisition is taken (see Starting a Measurement on page 273 for detai

Page 310

Managing Results370 FTB-5240S/5240BPManaging EDFA Test ResultsResults TabThe results table shows channel results for the active trace. The results for

Page 311

Managing ResultsOptical Spectrum Analyzer 371Managing EDFA Test Results³ Noise Figure: indicates the EDFA’s noise figure measured for the selected cha

Page 312 - Global Results Tab

Preparing Your Optical Spectrum Analyzer for a Test30 FTB-5240S/5240BPPerforming User Calibration6. To verify that the calibration changes are properl

Page 313 - Optical Spectrum Analyzer 305

Managing Results372 FTB-5240S/5240BPManaging EDFA Test ResultsMarkers TabYou can use markers to perform manual measurements and verification directly

Page 314 - Re-analyzing

Managing ResultsOptical Spectrum Analyzer 373Managing EDFA Test ResultsTo define a marker:1. From the main window, press the Markers tab.2. Set the ma

Page 315

Managing Results374 FTB-5240S/5240BPManaging EDFA Test ResultsYou can also move the markers directly on the EDFA Graph tab. Drag the marker to the des

Page 316 - 308 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 375Managing EDFA Test ResultsTrace Info. TabAfter acquiring a trace, you might want to view details about th

Page 317

Managing Results376 FTB-5240S/5240BPManaging EDFA Test Results3. Select the General tab.4. Edit the general information as required.Note: If you want

Page 318 - 2. Press Global Parameters

Managing ResultsOptical Spectrum Analyzer 377Managing EDFA Test ResultsTo edit trace information:1. From the main window, select the Trace Info. tab.2

Page 319

Managing Results378 FTB-5240S/5240BPManaging EDFA Test ResultsTo edit comments:1. From the main window, select the Trace Info. tab.2. Press Trace Iden

Page 320 - 312 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 379Managing EDFA Test ResultsChannel Results TabWhen you select a row from the Results tab, the Channel Resu

Page 321

Managing Results380 FTB-5240S/5240BPManaging EDFA Test Results2. Select a row from the Results tab to view the channel results for the selected channe

Page 322

Managing ResultsOptical Spectrum Analyzer 381Managing EDFA Test Results³ Channel index: indicates the channel number. Channel 1 corresponds to the fir

Page 323

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 31Performing User CalibrationTo clear user correction factors:1. From the

Page 324 - 316 FTB-5240S/5240BP

Managing Results382 FTB-5240S/5240BPManaging EDFA Test ResultsYou can include selected channels results in the global results calculation. To include

Page 325

Managing ResultsOptical Spectrum Analyzer 383Managing EDFA Test Results

Page 326

Managing Results384 FTB-5240S/5240BPManaging EDFA Test Results3. Select the Include in calculation option.

Page 327 - Optical Spectrum Analyzer 319

Managing ResultsOptical Spectrum Analyzer 385Managing EDFA Test ResultsTo modify channel results parameters:1. From the main window, select the Channe

Page 328 - Managing Drift Test Results

Managing Results386 FTB-5240S/5240BPManaging EDFA Test Results

Page 329 - ³ Signal power

Managing ResultsOptical Spectrum Analyzer 387Managing EDFA Test Results3. Edit the values as explained below:³ Channel Center (nm or THz): indicates t

Page 330

Managing Results388 FTB-5240S/5240BPManaging EDFA Test ResultsGlobal Results TabThe application allows you to view the global results of the current m

Page 331

Managing ResultsOptical Spectrum Analyzer 389Managing EDFA Test Results³ Average gain: indicates the sum of the gain of all detected channels in the c

Page 332 - Channel Graph Tab

Managing Results390 FTB-5240S/5240BPManaging EDFA Test ResultsChanging Active Trace Analysis ParametersThe global parameters modifies the active trace

Page 333 - Channel History Tab

Managing ResultsOptical Spectrum Analyzer 391Managing EDFA Test Results3. Under Default channel settings, modify the following parameters as needed: ³

Page 334

Contentsiv FTB-5240S/5240BP9 Setting Up the Instrument in EDFA Mode ...209Defining Preferences ...

Page 335

Preparing Your Optical Spectrum Analyzer for a Test32 FTB-5240S/5240BPPerforming User Calibration2. Select the Calibration tab.

Page 336 - WDM Graph Tab

Managing Results392 FTB-5240S/5240BPManaging EDFA Test Results4. Under Global analysis parameters, modify the following parameters as needed:³ Peak de

Page 337 - Optical Spectrum Analyzer 329

Managing ResultsOptical Spectrum Analyzer 393Managing EDFA Test Results³ Input power offset (dB): indicates the offset value applied on the input powe

Page 338

Managing Results394 FTB-5240S/5240BPManaging EDFA Test Results³ Output power offset (dB): indicates the offset value applied on the output power. You

Page 339 - Optical Spectrum Analyzer 331

Managing ResultsOptical Spectrum Analyzer 395Managing EDFA Test ResultsSwapping EDFA TracesThe Swap Trace feature allows you to swap EDFA input and ou

Page 340

Managing Results396 FTB-5240S/5240BPUsing Zoom ControlsUsing Zoom ControlsUse the zoom controls to change the scale of the trace display.You can zoom

Page 341

Managing ResultsOptical Spectrum Analyzer 397Using Zoom Controls³ You can zoom in or out on the center of the portion of the graph that is displayed b

Page 342 - 334 FTB-5240S/5240BP

Managing Results398 FTB-5240S/5240BPManaging Trace FilesManaging Trace FilesThe application allows you to manage the trace files for all test modes. Y

Page 343

Managing ResultsOptical Spectrum Analyzer 399Managing Trace Files2. Press New.

Page 344

Managing Results400 FTB-5240S/5240BPManaging Trace Files3. If you had already acquired (but not saved) a trace, a warning window appears, asking you i

Page 345 - Optical Spectrum Analyzer 337

Managing ResultsOptical Spectrum Analyzer 401Managing Trace FilesOpening Trace FilesOpening a trace file saves you time as you do not need to perform

Page 346

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 33Performing User Calibration3. Press Clear user correction factors, to l

Page 347 - Optical Spectrum Analyzer 339

Managing Results402 FTB-5240S/5240BPManaging Trace FilesOpening ST, EDFA and Legacy OSA trace files in WDM modeYour application allows you to open dif

Page 348 - Managing DFB Test Results

Managing ResultsOptical Spectrum Analyzer 403Managing Trace FilesIf the file contains two traces (Input and Output), then the application allows you t

Page 349 - Optical Spectrum Analyzer 341

Managing Results404 FTB-5240S/5240BPManaging Trace FilesOpening WDM trace file in DFB modeYour application allows you to open WDM file type in DFB mod

Page 350

Managing ResultsOptical Spectrum Analyzer 405Managing Trace FilesOpening WDM trace file in ST modeYour application allows you to open WDM file type in

Page 351 - Optical Spectrum Analyzer 343

Managing Results406 FTB-5240S/5240BPManaging Trace Files

Page 352

Managing ResultsOptical Spectrum Analyzer 407Managing Trace FilesOpening WDM trace file in EDFA modeYour application allows you to open WDM file type

Page 353 - Optical Spectrum Analyzer 345

Managing Results408 FTB-5240S/5240BPManaging Trace FilesAn example of WDM trace file imported in EDFA mode is shown below.

Page 354

Managing ResultsOptical Spectrum Analyzer 409Managing Trace FilesSaving Trace FilesThe application allows you to save the trace files for future refer

Page 355

Managing Results410 FTB-5240S/5240BPGenerating ReportsGenerating ReportsAfter performing any acquisition, the application allows you to generate a rep

Page 356 - 348 FTB-5240S/5240BP

Managing ResultsOptical Spectrum Analyzer 411Generating Reports2. Press Report.3. In the Save As window, enter a file name.4. From the Save as type li

Page 357

Preparing Your Optical Spectrum Analyzer for a Test34 FTB-5240S/5240BPPerforming User CalibrationTo save a user calibration:1. From the Main Menu, pre

Page 358 - ST Graph Tab

Managing Results412 FTB-5240S/5240BPGenerating ReportsFrom the Save as type list, select txt files (*.txt) to save the report in .txt format (if the f

Page 359 - Optical Spectrum Analyzer 351

Managing ResultsOptical Spectrum Analyzer 413Generating Reports5. Press Save. The report will be added to the Reports folder. You can change the locat

Page 360 - or frequency (nm or Thz):

Managing Results414 FTB-5240S/5240BPGenerating ReportsA sample WDM .txt report is shown below:

Page 361

Managing ResultsOptical Spectrum Analyzer 415Generating ReportsA sample Drift .html report is shown below:

Page 362

A sample Drift .txt report is shown below:

Page 363 - Optical Spectrum Analyzer 355

Managing ResultsOptical Spectrum Analyzer 417Generating ReportsA sample DFB .html report is shown below:

Page 364

Managing Results418 FTB-5240S/5240BPGenerating ReportsA sample Spectral Transmittance .html report is shown below:

Page 365 - Optical Spectrum Analyzer 357

Managing ResultsOptical Spectrum Analyzer 419Generating ReportsA sample EDFA .html report is shown below:

Page 367 - Optical Spectrum Analyzer 359

Optical Spectrum Analyzer 42112 MaintenanceTo help ensure long, trouble-free operation:³ Always inspect fiber-optic connectors before using them and c

Page 368

Preparing Your Optical Spectrum Analyzer for a TestOptical Spectrum Analyzer 35Performing User Calibration2. Select the Calibration tab.

Page 369 - Optical Spectrum Analyzer 361

Maintenance422 FTB-5240S/5240BPCleaning EUI ConnectorsCleaning EUI ConnectorsRegular cleaning of EUI connectors will help maintain optimum performance

Page 370

MaintenanceOptical Spectrum Analyzer 423Cleaning EUI Connectors5. Repeat steps 3 to 4 with a dry cleaning tip.Note: Make sure you don’t touch the soft

Page 371

Maintenance424 FTB-5240S/5240BPRecalibrating the UnitRecalibrating the UnitManufacturing and service center calibrations are based on the ISO/IEC 1702

Page 372

MaintenanceOptical Spectrum Analyzer 425Recycling and Disposal (Applies to European Union Only)Recycling and Disposal (Applies to European Union Only)

Page 374

Optical Spectrum Analyzer 42713 TroubleshootingViewing Online DocumentationA PDF version of the FTB-5240S/5240BP Optical Spectrum Analyzer user guide

Page 375 - Optical Spectrum Analyzer 367

Troubleshooting428 FTB-5240S/5240BPAboutAbout You can view the product version details and technical support information in this window.To view the in

Page 376 - Managing EDFA Test Results

TroubleshootingOptical Spectrum Analyzer 429About2. Press About.

Page 377 - Optical Spectrum Analyzer 369

Troubleshooting430 FTB-5240S/5240BPTransportationTransportationMaintain a temperature range within specifications when transporting the unit. Transpor

Page 378

Optical Spectrum Analyzer 43114 WarrantyGeneral InformationEXFO Inc. (EXFO) warrants this equipment against defects in material and workmanship for a

Page 379

Preparing Your Optical Spectrum Analyzer for a Test36 FTB-5240S/5240BPPerforming User Calibration3. Press Save Factors, to save the modified user cali

Page 380

Warranty432 FTB-5240S/5240BPLiabilityLiabilityEXFO shall not be liable for damages resulting from the use of the product, nor shall be responsible for

Page 381 - Optical Spectrum Analyzer 373

WarrantyOptical Spectrum Analyzer 433ExclusionsExclusionsEXFO reserves the right to make changes in the design or construction of any of its products

Page 382 - 374 FTB-5240S/5240BP

Warranty434 FTB-5240S/5240BPService and RepairsService and RepairsEXFO commits to providing product service and repair for five years following the da

Page 383

WarrantyOptical Spectrum Analyzer 435EXFO Service Centers WorldwideEXFO Service Centers WorldwideIf your product requires servicing, contact your near

Page 385

Optical Spectrum Analyzer 437A Technical SpecificationsIMPORTANTThe following technical specifications can change without notice. The information pres

Page 387

Optical Spectrum Analyzer 439B SCPI Command ReferenceThis appendix presents detailed information on the commands and queries supplied with your FTB-52

Page 388

SCPI Command Reference440 FTB-5240S/5240BPQuick Reference Command TreeQuick Reference Command TreeCommand Parameter(s) P.ABORt 443CALCulate[1..n]BWAT

Page 389 - Optical Spectrum Analyzer 381

SCPI Command ReferenceOptical Spectrum Analyzer 441Quick Reference Command TreeLEVel? 466POWer OFFSet [OUTPut] <Offset> 467[OUTPut]? 468RBW <

Page 390 - 382 FTB-5240S/5240BP

Optical Spectrum Analyzer 375 Setting Up the Instrument in WDM ModeBefore performing a spectral analysis in the WDM mode, you must set up your OSA mod

Page 391

SCPI Command Reference442 FTB-5240S/5240BPQuick Reference Command TreeSTORe CONFiguratio <Filename> 492OVERwrite <Overwrite> 493OVERwrite?

Page 392

SCPI Command ReferenceOptical Spectrum Analyzer 443Product-Specific Commands—DescriptionProduct-Specific Commands—Description:ABORtDescription This co

Page 393

SCPI Command Reference444 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:BWATDescription This command sets the power used to ca

Page 394

SCPI Command ReferenceOptical Spectrum Analyzer 445Product-Specific Commands—Description:CALCulate[1..n]:BWAT?Description This query returns the power

Page 395

SCPI Command Reference446 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel?Description This query returns available infor

Page 396

SCPI Command ReferenceOptical Spectrum Analyzer 447Product-Specific Commands—DescriptionB=Channel Center (always in m) <NR3 NUMERIC RESPONSE DATA&g

Page 397

SCPI Command Reference448 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel:ADDDescription Adds a new channel to list of c

Page 398 - 390 FTB-5240S/5240BP

SCPI Command ReferenceOptical Spectrum Analyzer 449Product-Specific Commands—DescriptionCentral wavelenght of the channel being added.Range is depende

Page 399

SCPI Command Reference450 FTB-5240S/5240BPProduct-Specific Commands—Description³ NoiseForOSNR: The program data syntax for the third parameter is defi

Page 400 - 392 FTB-5240S/5240BP

SCPI Command ReferenceOptical Spectrum Analyzer 451Product-Specific Commands—Description:CALCulate[1..n]:CHANnel:COUNt?Description This query returns

Page 401

Setting Up the Instrument in WDM Mode38 FTB-5240S/5240BP³ The most efficient way to setup the instrument is to use one of the favorites configurations

Page 402

SCPI Command Reference452 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault:NFOSnrDescription This command sets th

Page 403 - Swapping EDFA Traces

SCPI Command ReferenceOptical Spectrum Analyzer 453Product-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault:NFOSnr?Description This query

Page 404 - Using Zoom Controls

SCPI Command Reference454 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault:SIGPowercalcDescription This command s

Page 405

SCPI Command ReferenceOptical Spectrum Analyzer 455Product-Specific Commands—DescriptionPEAKSIGNAL, selects Peak signal power method to calculate sign

Page 406 - Managing Trace Files

SCPI Command Reference456 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault:SIGPowercalc?Description This query re

Page 407 - 2. Press New

SCPI Command ReferenceOptical Spectrum Analyzer 457Product-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault[:STATe]Description This comma

Page 408 - 400 FTB-5240S/5240BP

SCPI Command Reference458 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault[:STATe]?Description This query returns

Page 409 - Opening Trace Files

SCPI Command ReferenceOptical Spectrum Analyzer 459Product-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault:WIDThDescription This command

Page 410

SCPI Command Reference460 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel:DEFault:WIDTh?Description This query returns t

Page 411 - Optical Spectrum Analyzer 403

SCPI Command ReferenceOptical Spectrum Analyzer 461Product-Specific Commands—Description:CALCulate[1..n]:CHANnel:DELeteDescription This command delete

Page 412 - ³ Trace identification

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 39Defining PreferencesDefining PreferencesThe preferences window allows you to set gene

Page 413

SCPI Command Reference462 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:CHANnel:DELete:ALLDescription This command deletes all

Page 414

SCPI Command ReferenceOptical Spectrum Analyzer 463Product-Specific Commands—Description:CALCulate[1..n][:IMMediate]:AUTODescription This command indi

Page 415 - Optical Spectrum Analyzer 407

SCPI Command Reference464 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n][:IMMediate]:AUTO?Description This query returns the cu

Page 416

SCPI Command ReferenceOptical Spectrum Analyzer 465Product-Specific Commands—Description:CALCulate[1..n]:PEAK[:DETEction]:LEVelDescription This comman

Page 417 - Saving Trace Files

SCPI Command Reference466 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:PEAK[:DETEction]:LEVel?Description This query returns

Page 418 - Generating Reports

SCPI Command ReferenceOptical Spectrum Analyzer 467Product-Specific Commands—Description:CALCulate[1..n]:POWer:OFFSet[:OUTPut]Description This command

Page 419

SCPI Command Reference468 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:POWer:OFFSet[:OUTPut]?Description This querry returns

Page 420

SCPI Command ReferenceOptical Spectrum Analyzer 469Product-Specific Commands—Description:CALCulate[1..n]:RBWDescription This command sets the RBW for

Page 421

SCPI Command Reference470 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:RBW?Description This query gets the value for RBW for

Page 422

SCPI Command ReferenceOptical Spectrum Analyzer 471Product-Specific Commands—Description:CALCulate[1..n]:RBW:INSTrumentDescription This command sets w

Page 423

Setting Up the Instrument in WDM Mode40 FTB-5240S/5240BPDefining Preferences2. Select the General tab.3. Define the general parameters as needed.4. Pr

Page 424

SCPI Command Reference472 FTB-5240S/5240BPProduct-Specific Commands—DescriptionThe <RbwInstrumentState> parameter is a boolean value that determ

Page 425

SCPI Command ReferenceOptical Spectrum Analyzer 473Product-Specific Commands—Description:CALCulate[1..n]:RBW:INSTrument?Description This query returns

Page 426

SCPI Command Reference474 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:WAVelength:OFFSet[:OUTPut]Description This command set

Page 427

SCPI Command ReferenceOptical Spectrum Analyzer 475Product-Specific Commands—Description:CALCulate[1..n]:WAVelength:OFFSet[:OUTPut]?Description This q

Page 428

SCPI Command Reference476 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:WDMDescription When used alone, this command triggers

Page 429 - 12 Maintenance

SCPI Command ReferenceOptical Spectrum Analyzer 477Product-Specific Commands—Description:CALCulate[1..n]:WDM:CHANnel?Description This query returns th

Page 430

SCPI Command Reference478 FTB-5240S/5240BPProduct-Specific Commands—DescriptionE = Noise (always in dBm) <NR3 NUMERIC RESPONSE DATA>F = Noise ty

Page 431

SCPI Command ReferenceOptical Spectrum Analyzer 479Product-Specific Commands—Description:CALCulate[1..n]:WDM:CHANnel:COUNt?Description This query retu

Page 432 - Recalibrating the Unit

SCPI Command Reference480 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:WDM:OSNR:AVERage?Description This query returns the av

Page 433 - Recycling and Disposal

SCPI Command ReferenceOptical Spectrum Analyzer 481Product-Specific Commands—Description:CALCulate[1..n]:WDM:OSNR:FLATness?Description This query retu

Page 434

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 41Defining PreferencesTo enter link and location information:1. From the Main Menu, pre

Page 435 - 13 Troubleshooting

SCPI Command Reference482 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:WDM:SIGPower:AVERage?Description This query returns th

Page 436 - Troubleshooting

SCPI Command ReferenceOptical Spectrum Analyzer 483Product-Specific Commands—Description:CALCulate[1..n]:WDM:SIGPower:FLATness?Description This query

Page 437 - 2. Press About

SCPI Command Reference484 FTB-5240S/5240BPProduct-Specific Commands—Description:CALCulate[1..n]:WDM:SIGPower:INTegrated?Description This query returns

Page 438 - Transportation

SCPI Command ReferenceOptical Spectrum Analyzer 485Product-Specific Commands—Description:CALibration:ZERO[:IMMediate]Description This command performs

Page 439 - 14 Warranty

SCPI Command Reference486 FTB-5240S/5240BPProduct-Specific Commands—Description:INITiate[1..n][:IMMediate]Description This command completes one full

Page 440 - Liability

SCPI Command ReferenceOptical Spectrum Analyzer 487Product-Specific Commands—Description:MMEMory[1..n]:LOAD:CONFiguratioDescription This command loads

Page 441

SCPI Command Reference488 FTB-5240S/5240BPProduct-Specific Commands—Description:MMEMory[1..n]:LOAD:OVERwriteDescription This command selects, whether

Page 442 - Service and Repairs

SCPI Command ReferenceOptical Spectrum Analyzer 489Product-Specific Commands—Description:MMEMory[1..n]:LOAD:OVERwrite?Description This query lets you

Page 443 - Optical Spectrum Analyzer 435

SCPI Command Reference490 FTB-5240S/5240BPProduct-Specific Commands—Description:MMEMory[1..n]:LOAD:TRACeDescription This command loads a trace.For thi

Page 444

SCPI Command ReferenceOptical Spectrum Analyzer 491Product-Specific Commands—DescriptionSelects Trace index of destination.TRCMAIN, selects Main Trace

Page 445 - A Technical Specifications

ContentsOptical Spectrum Analyzer vA Technical Specifications ... 437B SCPI C

Page 446

Setting Up the Instrument in WDM Mode42 FTB-5240S/5240BPDefining Preferences3. Under System and link information, define the following parameters as n

Page 447 - B SCPI Command Reference

SCPI Command Reference492 FTB-5240S/5240BPProduct-Specific Commands—Description:MMEMory[1..n]:STORe:CONFiguratioDescription This command saves a confi

Page 448 - Quick Reference Command Tree

SCPI Command ReferenceOptical Spectrum Analyzer 493Product-Specific Commands—Description:MMEMory[1..n]:STORe:OVERwriteDescription This command sets, w

Page 449 - SCPI Command Reference

SCPI Command Reference494 FTB-5240S/5240BPProduct-Specific Commands—Description:MMEMory[1..n]:STORe:OVERwrite?Description This query lets you know whe

Page 450

SCPI Command ReferenceOptical Spectrum Analyzer 495Product-Specific Commands—Description:MMEMory[1..n]:STORe:TRACeDescription This command saves trace

Page 451 - Optical Spectrum Analyzer 443

SCPI Command Reference496 FTB-5240S/5240BPProduct-Specific Commands—Description:SENSe[1..n]:AVERage:COUNtDescription Sets the number of scans that wil

Page 452 - :CALCulate[1..n]:BWAT

SCPI Command ReferenceOptical Spectrum Analyzer 497Product-Specific Commands—Description:SENSe[1..n]:AVERage:COUNt?Description This query returns numb

Page 453 - :CALCulate[1..n]:BWAT?

SCPI Command Reference498 FTB-5240S/5240BPProduct-Specific Commands—Description:SENSe[1..n]:AVERage:INBand:COUNtDescription This command sets the numb

Page 454 - :CALCulate[1..n]:CHANnel?

SCPI Command ReferenceOptical Spectrum Analyzer 499Product-Specific Commands—Description:SENSe[1..n]:AVERage:INBand:COUNt?Description This query retur

Page 455

SCPI Command Reference500 FTB-5240S/5240BPProduct-Specific Commands—Description:SENSe[1..n]:AVERage:TYPEDescription This command sets the acquisition

Page 456 - :CALCulate[1..n]:CHANnel:ADD

SCPI Command ReferenceOptical Spectrum Analyzer 501Product-Specific Commands—Description:SENSe[1..n]:AVERage:TYPE?Description This query returns the a

Page 457

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 43Defining PreferencesTo enter comments:1. From the Main Menu, press Preferences.

Page 458

SCPI Command Reference502 FTB-5240S/5240BPProduct-Specific Commands—Description:SENSe[1..n]:WAVelength[:RANGe]:STARtDescription This command sets the

Page 459 - Optical Spectrum Analyzer 451

SCPI Command ReferenceOptical Spectrum Analyzer 503Product-Specific Commands—DescriptionExample(s) SENS:WAV:RANG:STAR 1520 NMNotes SENS:WAV:RANG:STOP

Page 460 - 452 FTB-5240S/5240BP

SCPI Command Reference504 FTB-5240S/5240BPProduct-Specific Commands—Description:SENSe[1..n]:WAVelength[:RANGe]:STARt?Description This query returns th

Page 461 - Optical Spectrum Analyzer 453

SCPI Command ReferenceOptical Spectrum Analyzer 505Product-Specific Commands—Description:SENSe[1..n]:WAVelength[:RANGe]:STOPDescription This command s

Page 462 - SIGPowercalc

SCPI Command Reference506 FTB-5240S/5240BPProduct-Specific Commands—DescriptionExample(s) SENS:WAV:RANG:STOP 1570 NMNotes SENS:WAV:RANG:STARt value wi

Page 463

SCPI Command ReferenceOptical Spectrum Analyzer 507Product-Specific Commands—Description:SENSe[1..n]:WAVelength[:RANGe]:STOP?Description This query re

Page 464 - SIGPowercalc?

SCPI Command Reference508 FTB-5240S/5240BPProduct-Specific Commands—Description:SNUMber?Description This query returns a value indicating the serial n

Page 465 - [:STATe]

SCPI Command ReferenceOptical Spectrum Analyzer 509Product-Specific Commands—Description:STATus?Description This query returns a value indicating the

Page 466 - [:STATe]?

SCPI Command Reference510 FTB-5240S/5240BPProduct-Specific Commands—DescriptionDEFECTIVE, means the module is defective andUNCONFIGURED, means the mod

Page 467 - ³ ChannelWidthUnit:

SCPI Command ReferenceOptical Spectrum Analyzer 511Product-Specific Commands—Description:TRACe[1..n][:DATA]?Description This query returns the points

Page 468 - 460 FTB-5240S/5240BP

Setting Up the Instrument in WDM Mode44 FTB-5240S/5240BPDefining Preferences2. Select the Comments tab. 3. Enter your comments for the current trace.

Page 469 - Optical Spectrum Analyzer 461

SCPI Command Reference512 FTB-5240S/5240BPProduct-Specific Commands—Description:TRACe[1..n][:DATA]:PREamble?Description This query returns a trace’s h

Page 470 - 462 FTB-5240S/5240BP

SCPI Command ReferenceOptical Spectrum Analyzer 513Product-Specific Commands—DescriptionA = Acquisition type <STRING RESPONSE DATA>B = Spectral

Page 471 - Optical Spectrum Analyzer 463

SCPI Command Reference514 FTB-5240S/5240BPProduct-Specific Commands—Description:TRACe[1..n]:INFormation?Description This query returns all available i

Page 472 - 464 FTB-5240S/5240BP

SCPI Command ReferenceOptical Spectrum Analyzer 515Product-Specific Commands—DescriptionA = Acquisition start date and time <STRING RESPONSE DATA&g

Page 473 - Optical Spectrum Analyzer 465

SCPI Command Reference516 FTB-5240S/5240BPProduct-Specific Commands—Description:TRACe[1..n]:POINts?Description This query returns the number of points

Page 474 - 466 FTB-5240S/5240BP

SCPI Command ReferenceOptical Spectrum Analyzer 517Product-Specific Commands—Description:UNIT[1..n]:SPECtrumDescription This command sets the spectrum

Page 475 - Parameter(s) Offset:

SCPI Command Reference518 FTB-5240S/5240BPProduct-Specific Commands—Description:UNIT[1..n]:SPECtrum?Description This query returns the current spectru

Page 476 - for global channel

Optical Spectrum Analyzer 519C Formulas Used with Your Optical Spectrum AnalyzerThe following formulas are used in the various tests available with yo

Page 477 - :CALCulate[1..n]:RBW

Formulas Used with Your Optical Spectrum Analyzer520 FTB-5240S/5240BPCentral Wavelength Calculation (Spectral Transmittance)Central Wavelength Calcula

Page 478 - :CALCulate[1..n]:RBW?

Formulas Used with Your Optical Spectrum AnalyzerOptical Spectrum Analyzer 521Bandwidth Calculation (Spectral Transmittance)Bandwidth Calculation (Spe

Page 479 - Optical Spectrum Analyzer 471

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 45Defining PreferencesDefining Display ParametersThe application allows you to set disp

Page 481 - Optical Spectrum Analyzer 473

IndexOptical Spectrum Analyzer 523IndexAacquisitionaveraging... 99, 159, 179, 205, 265real-time ... 99, 179, 205, 265

Page 482 - Range: [-99.999 … 99.999] nm

Index524 FTB-5240S/5240BPFFabry-Perot mode spacing... 343fiber ends, cleaning... 15fiber ID ..

Page 483 - [:OUTPut]?

IndexOptical Spectrum Analyzer 525wavelength... 64, 129online user guide... 427op

Page 484 - :CALCulate[1..n]:WDM

Index526 FTB-5240S/5240BPspecifications, product ... 437statusfail ...

Page 485 - :CALCulate[1..n]:WDM:CHANnel?

NOTICE抩⛙CHINESE REGULATION ON RESTRICTION OF HAZARDOUS SUBSTANCES₼⦌␂ℝ☀⹂䓸德棟Ⓟ䤓屓⸩NAMES AND CONTENTS OF THE TOXIC OR HAZARDOUS SUBSTANCES OR ELEMENTS CONT

Page 486

MARKING REQUIREMENTS㪖㽷尐㻑Productℶ❐Environmental protection use period (years)䘾⬒≬㔳∎䞷㦮棟 ( ㄃ )Logo㪖㉦This Exfo product㦻 EXFO ℶ❐10Batterya䟄㻯aa. If applicabl

Page 487 - Optical Spectrum Analyzer 479

www.EXFO.com · [email protected]/N: 1059976CORPORATE HEADQUARTERS 400 Godin Avenue Quebec (Quebec) G1M 2K2 CANADATel.: 1 418 683-0211 · Fax: 1 418 683-

Page 488 - 480 FTB-5240S/5240BP

Setting Up the Instrument in WDM Mode46 FTB-5240S/5240BPDefining Preferences2. Select the Display tab.

Page 489 - Optical Spectrum Analyzer 481

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 47Defining Preferences3. Select the spectral unit you want to work with. With the nanom

Page 490 - AVERage?

Setting Up the Instrument in WDM Mode48 FTB-5240S/5240BPDefining Preferences4. Select the label that will appear on the peaks in the graph.

Page 491 - FLATness?

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 49Defining PreferencesNote: The channel name and channel number cannot be shown at the

Page 492 - INTegrated?

Setting Up the Instrument in WDM Mode50 FTB-5240S/5240BPDefining Preferences5. Select Show to display the empty channels from the channel list in the

Page 493 - :CALibration:ZERO[:IMMediate]

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 51Defining Preferences7. Select the background color scheme for Graph as desired. 8. Pr

Page 494 - :INITiate[1..n][:IMMediate]

Certification Informationvi FTB-5240S/5240BPCertification InformationF.C.C. InformationElectronic test equipment is exempt from Part 15 compliance (FC

Page 495 - Optical Spectrum Analyzer 487

Setting Up the Instrument in WDM Mode52 FTB-5240S/5240BPDefining PreferencesCustomizing WDM Results TableIt is possible to select which results you wo

Page 496 - :MMEMory[1..n]:LOAD:OVERwrite

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 53Defining Preferences2. Select the WDM Results tab. 3. Select which parameters you wan

Page 497 - Optical Spectrum Analyzer 489

Setting Up the Instrument in WDM Mode54 FTB-5240S/5240BPDefining Preferences4. Press Up or Down to change the order in which the columns will appear i

Page 498 - :MMEMory[1..n]:LOAD:TRACe

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 55Defining PreferencesCustomizing WDM File NameThe application shall provide a way to d

Page 499

Setting Up the Instrument in WDM Mode56 FTB-5240S/5240BPDefining Preferences2. Select the File Name tab. 3. Select which parameters you want to includ

Page 500 - 492 FTB-5240S/5240BP

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 57Defining Preferences4. Press Up or Down to change the order in which the field values

Page 501 - Optical Spectrum Analyzer 493

Setting Up the Instrument in WDM Mode58 FTB-5240S/5240BPSetting Up WDM Analysis ParametersSetting Up WDM Analysis ParametersThis section presents the

Page 502 - 494 FTB-5240S/5240BP

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 59Setting Up WDM Analysis ParametersDefining General SettingsThe general analysis param

Page 503 - :MMEMory[1..n]:STORe:TRACe

Setting Up the Instrument in WDM Mode60 FTB-5240S/5240BPSetting Up WDM Analysis Parameters2. Select the General tab.

Page 504 - :SENSe[1..n]:AVERage:COUNt

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 61Setting Up WDM Analysis Parameters3. Under Default channel settings, define the follo

Page 505 - :SENSe[1..n]:AVERage:COUNt?

Certification InformationOptical Spectrum Analyzer vii Application of Council Directive(s): 2006/95/EC - The Low Voltage Directive 2004/108/EC -

Page 506 - 498 FTB-5240S/5240BP

Setting Up the Instrument in WDM Mode62 FTB-5240S/5240BPSetting Up WDM Analysis Parameters³ Channel width (GHz or nm): indicates the limit inside whic

Page 507 - Optical Spectrum Analyzer 499

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 63Setting Up WDM Analysis Parameters³ Noise for OSNR: indicates which calculation metho

Page 508 - :SENSe[1..n]:AVERage:TYPE

Setting Up the Instrument in WDM Mode64 FTB-5240S/5240BPSetting Up WDM Analysis Parameters³ Peak detection level (dBm): indicates the minimum power le

Page 509 - :SENSe[1..n]:AVERage:TYPE?

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 65Setting Up WDM Analysis Parameters5. If you want to apply the above settings to the a

Page 510 - 502 FTB-5240S/5240BP

Setting Up the Instrument in WDM Mode66 FTB-5240S/5240BPSetting Up WDM Analysis ParametersYou can set your pass/fail threshold limits in different way

Page 511 - Optical Spectrum Analyzer 503

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 67Setting Up WDM Analysis ParametersDefining Global ThresholdsGlobal thresholds will be

Page 512 - 504 FTB-5240S/5240BP

Setting Up the Instrument in WDM Mode68 FTB-5240S/5240BPSetting Up WDM Analysis ParametersTo define global thresholds:1. From the Main Menu, press Ana

Page 513 - Optical Spectrum Analyzer 505

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 69Setting Up WDM Analysis Parameters3. Select the Activate all thresholds option, to ma

Page 514 - 506 FTB-5240S/5240BP

Setting Up the Instrument in WDM Mode70 FTB-5240S/5240BPSetting Up WDM Analysis Parameters³ Empty channel count: Select this option to obtain the numb

Page 515 - Optical Spectrum Analyzer 507

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 71Setting Up WDM Analysis ParametersDefining Default ThresholdsDefault thresholds will

Page 517 - :STATus?

Setting Up the Instrument in WDM Mode72 FTB-5240S/5240BPSetting Up WDM Analysis Parameters2. Select the Default Thresholds tab. 3. Enter values in the

Page 518

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 73Setting Up WDM Analysis Parameters4. If you want to apply the above settings to the a

Page 519 - :TRACe[1..n][:DATA]?

Setting Up the Instrument in WDM Mode74 FTB-5240S/5240BPSetting Up WDM Analysis ParametersManaging ChannelsTesting DWDM systems involves characterizin

Page 520 - :TRACe[1..n][:DATA]:PREamble?

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 75Setting Up WDM Analysis Parameters2. Select the Channels tab. 3. By default, the chan

Page 521

Setting Up the Instrument in WDM Mode76 FTB-5240S/5240BPSetting Up WDM Analysis Parameters³ Stop range (nm or Thz): indicates the ending range of the

Page 522 - :TRACe[1..n]:INFormation?

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 77Setting Up WDM Analysis Parameters6. If you want to apply the above settings to the a

Page 523

Setting Up the Instrument in WDM Mode78 FTB-5240S/5240BPSetting Up WDM Analysis ParametersTo edit the parameters of a specific channel:1. From the Mai

Page 524 - :TRACe[1..n]:POINts?

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 79Setting Up WDM Analysis Parameters2. Select the Channels tab. 3. Select the channel o

Page 525 - :UNIT[1..n]:SPECtrum

Setting Up the Instrument in WDM Mode80 FTB-5240S/5240BPSetting Up WDM Analysis Parameters4. Press Edit Selection.

Page 526 - :UNIT[1..n]:SPECtrum?

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 81Setting Up WDM Analysis Parameters5. Modify the settings as needed. For more informat

Page 527 - C Formulas Used with Your

Optical Spectrum Analyzer 11 Introducing the FTB-5240S/5240BP Optical Spectrum AnalyzerThe FTB-5240S/5240BP Optical Spectrum Analyzer (OSA) is designe

Page 528 - Transmittance)

Setting Up the Instrument in WDM Mode82 FTB-5240S/5240BPSetting Up WDM Analysis Parameters6. Press OK to return to the Channels tab, which now contain

Page 529

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 83Setting Up WDM Analysis ParametersTo add current peaks:Note: You can add current peak

Page 530

Setting Up the Instrument in WDM Mode84 FTB-5240S/5240BPSetting Up WDM Analysis Parameters2. Select the Channels tab. 3. Press Import from Trace. All

Page 531 - Optical Spectrum Analyzer 523

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 85Setting Up WDM Analysis Parameters4. If you want to apply the above settings to the a

Page 532

Setting Up the Instrument in WDM Mode86 FTB-5240S/5240BPSetting Up WDM Analysis ParametersManaging FavoritesFavorites are configurations files that co

Page 533 - Optical Spectrum Analyzer 525

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 87Setting Up WDM Analysis Parameters2. Select the Favorites tab.

Page 534

Setting Up the Instrument in WDM Mode88 FTB-5240S/5240BPSetting Up WDM Analysis Parameters3. To apply the settings from a favorite file to the current

Page 535 - ₼⦌␂ℝ☀⹂䓸德棟Ⓟ䤓屓⸩

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 89Setting Up WDM Analysis Parameters4. If you want to apply the above settings to the a

Page 536 - MARKING REQUIREMENTS

Setting Up the Instrument in WDM Mode90 FTB-5240S/5240BPSetting Up WDM Analysis ParametersTo save a test configuration:1. From the Main Menu, press An

Page 537

Setting Up the Instrument in WDM ModeOptical Spectrum Analyzer 91Setting Up WDM Analysis Parameters2. Select the Favorites tab.3. To save an analysis